RU

SEM-6M-2 Channel Type Secondary Electron Multiplier

The secondary electron multiplier of the channel type with an extended dynamic range SEM-6M-2 is designed for recording charged particles, ultraviolet and X-ray radiation.

The device body is made of PEEK plastic, which is widely used for work in a vacuum. The dimensions of the developed device are completely identical to the dimensions of foreign analogues, which allows it to be used for equipment originally developed for the form factor of foreign-made devices (for example, Photonis Magnum 5900, DeTech 2300).

Specification

Scheme

Basic performance

Наименование параметраНорма
Gain at supply voltage Usupply ≤ 2000 V1×107,  min 
Gain at supply voltage Usupply ≤ 2300 V1×108,  min 
Dark count rate, counts/s0.5, max
Channel strip current at Usupply  = 3000 V, µA70, min
Linear output current, µA10% of strip current, min
Pulse height resolution at count rate of 1×103 to 5×103 count/s and Usupply corresponding to gain of 1×108 min     0.8, max