Detector Type MCPs

Detector type MCPs are used to amplify weak signals and determine position parameters of particles in a variety of image detectors, as part of photoelectronic and X-ray vacuum devices, field-ion microscopes, mass-spectrometers of various types.

Photomultiplier tubes (PMT) with detector type MCPs have the advantage of fast response and gain, increased surface area, low power consumption, and stable operation under the exposure to magnetic fields.

Applications of detector type MCPs

Specification

Decryption of the name

1) Rimmed MCP — rimmed microchannel plate
2) 18 — active area diameter, mm
3) 10 — channel diameter, mm
4) А — with increased bias angle

Schematic diagram

Microchannel plates with 46 mm, 50 mm, 56 mm, 87 mm diameters
Parameter, unit Microchannel plate type
MCP 46-12 MCP 50-10 Rimmed MCP 50-10 MCP 50-15 MCP 56-15 Rimmed MCP 56-15 MCP 70-15 MCP 87-12
Nominal value
 Geometrics
 Outer diameter D, mm 45.7–46.0 49.5 49.8–50.0 56 69.5–70.0 86.3–86.8
Active area diameter, DA, mm, min 4­4 48.5 41.5 47.5 54 48 68.0 77.5
Thickness L, mm 0.46–0.50 0.5 0.75–0.80 0.7–0.8 0.69–0.75
Channel bias angle θ, degree 5°–9° 6°–9°
11°–13°
7°–10°     9°–12°
Structural parameters
Channel diameter, μm 11.1–12.5 10.0–10.6 14.5–15.5 11.5–12.5
Channel pitch, µm 15 max от 12.2 до 12.5 19 max 15
Contact electrode material Chromium
Electrical ratings
Rated voltage at gain of 104, V, max 1100 1200
MCP resistance, ×108 Ohm 0,5–10 0.25–2.5 0.5–10 0.5–2.5 0.05–0.5