The detector with a voltage divider on the printed circuit board is designed for the detection of charged particle fluxes as part of a mass spectrometer.

The board on which the MCP detector is mounted consists of a detector voltage divider, components of the power supply circuit and a coaxial connector for the signal readout. The assembly of the detector on the board allows to reduce the influence of spurious inductances, and as a result, the output pulse of the MCP detector acquires a classical waveform without “ripple” with improved timing characteristics. The compact design with 2 pins for the power supply and a coaxial connector for the signal acquisition is more appropriate to the application and provides easy installation and replacement of the detector when needed.




Specification
Basic parameters
Parameter, unit | Value | |
---|---|---|
min | max | |
MCP active area diameter, mm | 25 | |
Supply voltage at 1×107 gain, V | 2400 | |
Single electron pulse height resolution, % | 135 | |
Dark count rate density, count/sec×cm2 | 3 | |
Average pulse width (FWHM), ns | 1.6 | |
Pulse leading edge rise time, ns | 0.75 | |
Product consumption current at supply voltage, μA | 250 |
Typical output waveform

The detector electrical connection diagram

R1 = 10 MΩ;
R2 = R3 = 1 MΩ;
С1 = С2 = 470 pF;
Usupply = -2.4 kV, max
1 – control electrode;
2 – MCP unit;
3 – anode
General view, overall, setting and mounting dimensions of MCP detector with 25 mm active area diameter



Lead designation | |
---|---|
1 | Supply voltage –2400 V, max |
2. 3 | Package |
SMA | Signal output |
А | Anode |