Detector with MCP 25-10 and open input

The detector is intended for the corpuscular and electromagnetic radiation registration in the scientific and industrial equipment. The detector consists of MCP-stack, metal–ceramic fittings and a metal collector. The operating principle is based on the charged particles input current registration and further amplification.

Detector's view
 
Dimensional outlines
Physical Characteristics

Parameter

Dimension

Value

MCP Channel Diameterum10
Bias Angledegree7; 13
Active Diametermm25
Maximum Detector Outer Diametermm54
Maximum Detector Heightmm13
Number of MCP-2
Electrical Characteristics

Parameter

Dimension

Value

Minimum gain-1×107
Maximum Single-electron Amplitude Resolution%130
Maximum Dark Pulses Count Rate Densitypulse/s-1×cm-23
MCP ResistanceΩ5×107÷2×108
Maximum Operating VoltageV2800
Vacuum ≤Pa (Torr)1,3×10-4 (1×10-6)
Circuit diagram